X-ray tomography in coordinate metrology

Perform measurement of the entire exterior and interior

PRACTICAL TIP PART 3

In coordinate metrology, tactile and optical sensors, as well as X-ray computed tomography, are predominantly used. X-ray tomography machines differ in terms of their basic design and components – such as the X-ray source, rotary axis and detector – as well as their software; consequently, they vary in their characteristics, and a fundamental understanding of these is helpful for their optimal use. Spatially extensive objects, including their internal structures, can be fully captured by metrology.

X-ray tomography makes use of the ability of X-rays to penetrate objects. X-rays emanate from a point-source and pass through the measuring object to reach the X-ray sensor. As they pass through the measuring object, some of the radiation is absorbed. The longer the path length within the measuring object, the less radiation emerges from the other side. Furthermore, the absorption also depends on the type of material. By shifting the rotary axis or the measuring object relative to the X-ray unit (source and sensor), the magnification – and thus the resolution – can be adjusted when capturing a measuring object.

To scan tomographically an object, several hundred to a few thousand such two-dimensional radiographic images are captured in succession – in the case of OnTheFly Tomography, whilst the rotary axis is in continuous motion – at various rotational positions of the measuring object. Mathematical methods are then used to calculate a volume model from these images, which fully describes the geometry and material distribution of the workpiece (Figure 1). To determine sizes from the volume data, the exact position of material transitions (for example, from metal to air) is calculated from the amplitudes of the surrounding voxels (volumetric pixels). By incorporating the amplitude information, the achievable resolution for edge location determination is significantly higher than that determined by the centre-to-centre distance of the voxel grid (subvoxelling).

To determine sizes from the measurement points generated in this way, geometric elements such as lines, cylinders or planes are calculated from groups of points. The points are usually selected with the aid of a CAD model. From the geometric elements determined in this way, the sizes are calculated by establishing relationships (e.g. the distance between two planes or the centres of two circles) and compared with the target values.

Perform measurement of the entire exterior and interior

Figure 1. Calculation of volume data by back-projection of filtered radiographic images: a) Object, b) X-ray path in the cross-sectional plane, c) Principle of stepwise back-projection and superposition, d) Reconstruction result with varying numbers of back-projections on the actual workpiece (© Werth)

X-ray tube influences resolution

The X-ray tubes used to generate X-rays are a core component of tomography measuring instruments. The X-rays are produced in an evacuated tube when a high-energy electron beam strikes a metal target. The energy of the X-rays produced depends on the voltage between the cathode and anode of the X-ray tube and on the target material. This is important when selecting an X-ray tube, as the radiation energy must be matched to the material of the workpiece to ensure optimum measurement results.

Generally speaking, X-ray tube targets are classified as either reflection targets or transmission targets. The difference between using reflection and transmission targets lies in the available radiation power – and consequently the measurement time – in conjunction with the minimum focal spot size that can be achieved.

With a reflection target, the X-rays are reflected by the target. The disadvantage of reflection targets is that small focal spot sizes can only be achieved at very low power and thus with a long measurement time (typically 5 µm at 5 W). In most cases, however, a higher maximum power is available – albeit with lower resolution – which is suitable for the measurement of large workpieces. Transmission targets are penetrated by the X-rays. Here, a small focal spot can be achieved even at medium power levels (typically 5 µm at 25 W), and most workpieces can be measured quickly with sufficient resolution.

Open-design X-ray tubes require regular maintenance, whilst closed design tubes must be completely replaced according to a few years. Transmission target tubes in a monoblock design combine the benefits of both designs, with a long maintenance interval of typically 12 months and a technically unlimited service life.

Software methods for universal use

In addition to the method described above for determining geometrical characteristics (sizes) with the aid of CAD data, there are a number of other evaluation methods (Figure 2). Geometric elements can be calculated from the point cloud without CAD data using an automatic segmentation function, starting from a reference point. During the 3D nominal-actual comparison, the software calculates the distances between the individual measurement points and the CAD surface and displays these using colour coding. Deviations of the actual geometry from the target geometry are thus immediately apparent. Data for correcting components such as injection moulding tools is also calculated directly. Further methods are used to determine drawing dimensions in 2D views and cross sections using image processing or contour analysis techniques. Other software tools are used for the automatic identification of voids or inclusions in the measuring object, or of burrs and chips.

With raster CT, the measuring range can be extended or the resolution for the entire workpiece increased by stitching together several tomographed sections of the workpiece. Eccentric ROI-CT enables high-resolution measurements and the linking of several parts of the workpiece.

Perform measurement of the entire exterior and interior

Figure 2. CAD model, CT volume data and point cloud with measured geometrical characteristics, color-coded deviation plot from a 3D nominal-actual comparison and 2D cross sections (© Werth)